Remote Control Commands
R&S
®
SMW200A
994User Manual 1175.6632.02 ─ 16
Manual operation: See "Carrier Table" on page 378
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:PHASe[:STARt] <Start>
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:POWer[:STARt] <Start>
Sets the power/pahse for the starting carrier. The power of the remaining carriers is
stepped up or down by the power specified with the [:SOURce<hw>]:BB:MCCW:
EDIT:CARRier:POWer:STEP command.
Parameters:
<Start> float
Range: -80 to 0
Increment: 0.01
*RST: 0
Example: see "Programming Examples" on page 983
Manual operation: See "Power Start" on page 377
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:PHASe:STEP
<Step>
For disabled optimization of the crest factor, sets the step width by which the start
phase of the carriers in the defined carrier range is incremented.
Parameters:
<Step> float
Range: -359.99 to 359.99
Increment: 0.01
*RST: 0
Example: See "Programming Examples" on page 983
Manual operation: See "Phase Step" on page 378
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:POWer:STEP <Step>
Sets the step width by which the starting power of the carriers in the defined carrier
range is incremented.
Parameters:
<Step> float
Range: -80 to 80
Increment: 0.01
*RST: 0
Example: See "Programming Examples" on page 983
Manual operation: See "Power Step" on page 378
SOURce Subsystem