Remote Control Commands
R&S
®
SMW200A
1010User Manual 1175.6632.02 ─ 16
Usage: Query only
Manual operation: See "Carrier + Noise Power, Signal + Noise Power (System
Bandwidth), Carrier + Interferer Power, Signal + Interferer Power
" on page 403
[:SOURce<hw>]:AWGN:POWer:SUM:PEP?
Queries the peak envelope power of the overall signal comprised of noise signal plus
useful signal.
Return values:
<Pep> float
Range: -145 to 20
Increment: 0.01
*RST: 0
Example: see Example "Generating an additive noise signal"
on page 1004
Usage: Query only
Manual operation: See "Carrier + Noise PEP, Signal + Noise PEP (Total Band-
width), Carrier + Interferer PEP, Signal + Interferer PEP"
on page 403
[:SOURce<hw>]:AWGN:FREQuency:TARGet <Target>
Sets the desired frequency of the sine wave.
Parameters:
<Target> float
Range: -40E6 to 40E6
Increment: 0.01
*RST: 0
Example: see Example "Generating a CW interferer signal" on page 1003
Manual operation: See "Target CW Frequency Offset" on page 397
[:SOURce<hw>]:AWGN:FREQuency:RESult?