Remote Control Commands
R&S
®
SMW200A
1097User Manual 1175.6632.02 ─ 16
Example:
POW:EMF:STAT 1
Activates voltage level display.
Manual operation: See " Display Level as Voltage of EMF " on page 660
[:SOURce<hw>]:POWer:LBEHaviour <Behaviour>
Selects the level behavior at the RF output over time.
Parameters:
<Behaviour> AUTO | UNINterrupted | MONotone | CVSWr | CPHase
UNINterrupted|MONotone
Do not use the uninterrupted level settings and strictly monotone
modes in combination with the high-quality optimization mode
(see [:SOURce<hw>]:BB:IMPairment:OPTimization:
MODE on page 1015)
CWSWr
Constant VSWR
CPHase
Continuous phase
*RST: AUTO
Example:
SOURce1:POWer:LBEHaviour AUTO
Manual operation: See " Setting Characteristics " on page 436
[:SOURce<hw>]:POWer:LIMit[:AMPLitude] <Amplitude>
Limits the maximum RF output level in CW and sweep mode.
It does not influence the "Level" display or the response to the query [:
SOURce<hw>]:POWer[:LEVel][:IMMediate][:AMPLitude].
Parameters:
<Amplitude> float
Range: depends on the installed options
Increment: 0.01
*RST: n.a. (factory preset: 30)
Example:
SOURce1:POWer:LIMit:AMPLitude 10
Manual operation: See " Limit " on page 438
[:SOURce<hw>]:POWer:LMODe <LevMode>
Sets the RF level mode.
SOURce Subsystem