Remote Control Commands
R&S
®
SMW200A
1101User Manual 1175.6632.02 ─ 16
Example:
// increasing the RF level with a step size of 2 dB
SOURce1:POWer:STEP:INCRement 2
SOURce1:POWer:STEP:MODE USER
SOURce1:POWer:LEVel:IMMediate:AMPLitude UP
Manual operation: See "Variation Active" on page 433
[:SOURce<hw>]:POWer:STEP[:INCRement] <Increment>
Specifies the step width in the appropriate path for POW:STEP:MODE USER.
To adjust the level step-by-step with this increment value, use the command POW UP,
or POW DOWN.
Note: The command also sets "Variation Step" in the manual control, that means the
user-defined step width for setting the level with the rotary knob or the UP/DOWN
arrow keys.
Parameters:
<Increment> float
Range: 0 to 100
Increment: 0.01
*RST: 1
Default unit: dB
Example: See [:SOURce<hw>]:POWer:STEP:MODE on page 1100.
Manual operation: See "Variation Step" on page 433
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:OFFSet <Offset>
Sets the level offset of a downstream instrument.
The level at the RF output is not changed.
To query the resulting level, as it is at the output of the downstream instrument, use the
command [:SOURce<hw>]:POWer[:LEVel][:IMMediate][:AMPLitude].
See Chapter 7.3.4.1, "Displayed RF Frequency and Level Values with Downstream
Instruments", on page 440.
Note: The level offset also affects the RF level sweep.
Parameters:
<Offset> float
Range: -100 to 100
Increment: 0.01
*RST: 0
Default unit: dB
Level offset is always expreced in dB; linear units (V, W, etc.)
are not supported
Example:
POWer:OFFSet 10
Sets the RF level offset to 10 dB
SOURce Subsystem