EasyManua.ls Logo

Rohde & Schwarz SMW200A - Page 1115

Rohde & Schwarz SMW200A
1312 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMW200A
1100User Manual 1175.6632.02 ─ 16
Parameters:
<Power> float
Level at the RF output, without level offset
Range: See data sheet
Increment: 0.01
Default unit: dBm
Example:
SOURce1:POWer:POWer 15
Sets the level at RF output
Manual operation: See "Amplitude" on page 435
[:SOURce<hw>]:POWer:STARt
<Start>
[:SOURce<hw>]:POWer:STOP <Stop>
Sets the RF start/stop level in sweep mode.
Parameters:
<Stop> float
Sets the setting range calculated as follows:
(Level_min + OFFSet) to (Level_max + OFFSet)
Where the values are set with the commands:
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:OFFSet
[:SOURce<hw>]:POWer:STARt
[:SOURce<hw>]:POWer:STOP
Range: Minimum level to maximum level
*RST: -30 (Start)/ -10 (Stop)
Default unit: dBm
Example: See Example "Setup an RF frequency or power sweep"
on page 1107
Manual operation: See " Start Level / Stop Level " on page 478
[:SOURce<hw>]:POWer:STEP:MODE <Mode>
Defines the type of step width to vary the RF output power step-by-step with the com-
mands POW UP or POW DOWN.
You can define the step mode for each path separately.
Parameters:
<Mode> DECimal | USER
DECimal
Increases or decreases the level in steps of ten.
USER
Increases or decreases the level in increments, determined with
the command [:SOURce<hw>]:POWer:STEP[:INCRement].
*RST: DECimal
SOURce Subsystem

Table of Contents

Related product manuals