Configuring the Internal Baseband Source
R&S
®
SMW200A
245User Manual 1175.6632.02 ─ 16
Simultaneous Signal Generation Start in All Basebands
The R&S SMW supports you to trigger all basebands simultaneous in the following
cases:
In system configuration with Coupled sources
See "Coupled or separate baseband sources" on page 128
In configuration that involves signal routing with signal addition, like:
Any MIMO configuration
See Chapter 3.2.2, "Using the System Configuration Capabilities", on page 127
Routing and summing of Baseband and/or Streams signals before and after the
"Fading" blocks
See Chapter 3.2.1, "Using the Signal Routing Settings and the Block Diagram",
on page 126
Routing and summing of Streams in the "I/Q Stream Mapper"
See "I/Q stream mapper" on page 129
In instruments equipped with 2xR&S SMW-B10, 1xR&S SMW-B14 and R&S SMW-
B13T.
In master-slave mode
See Chapter 9.3.1, "Connecting Multiple Instruments in Master-Slave Mode",
on page 608
To enable simultaneous signal generation in all basebands in any of these cases, the
R&S SMW couples automatically the trigger settings in all available basebands. It is
sufficient to adjust the trigger settings, like trigger source and trigger mode in one of
the basebands; the R&S SMW applies them automatically to all other. An arm or a
restart trigger event applies to all basebands, too. You can still apply different delay to
each of the triggers individually.