Configuring the Internal Baseband Source
R&S
®
SMW200A
346User Manual 1175.6632.02 ─ 16
3. In the "ARB: Multi Segment > Level/Clock/Marker" dialog, adjust the "Level",
"Clock", and "Segment Marker" settings.
4. Select the "ARB > Multi Segment > General > Output File" and enter file name.
5. Select "ARB > Multi Segment > General > Save List" to store the configuration file.
6. Select "ARB > Multi Segment > General > Create" or "Create and Load" to create
and load the multi-segment waveform.
7. Select "ARB > Multi Segment > General > State > On".
Tip: Try out the built-in tutorial.
The following description emphasizes on the settings required to fulfill the particular
task or achieve the desired performance; standard basic settings are not discussed.
All provided examples use the example segment sequence as shown on Figure 4-22.
Figure 4-22: Example: multi-segment sequence with marker signals
See how to:
●
"To apply a manual segment switch to the output segments in the required order"
on page 346
●
"To generate a multi-segment signal without signal interruptions" on page 348
●
"To enable an external triggering for high-speed switch over" on page 348
●
"To configure a multi-segment sequence using the play list function" on page 349
●
"To add marker signals" on page 350
To apply a manual segment switch to the output segments in the required order
When testing DUTs using different test signals, it may be necessary to switch dynami-
cally and flexibly between any of the test signals, for instance depending on the out-
come of the previous test. The following settings enable flexible dynamic switch over
between the segments in a multi-segment waveform.
Generating Multi-Segment Waveform Files