Configuring the Signal in the RF Domain
R&S
®
SMW200A
462User Manual 1175.6632.02 ─ 16
●
Switches automatically to the next step when the Dwell time has elapsed.
●
Stops signal generation at the set end value and waits for the subsequent trigger
event.
●
Trigger sources:
– The "Execute Single" function.
– The corresponding remote control command.
– An externally applied trigger signal.
●
"State = Off" stops the signal generation in sweep or list mode.
Table 7-4: Cross-reference between manual and remote control in Single / Extern Single modes
(Sweep/List)
Manual control mode:
"Single / Extern Single"
Remote commands
Rohde & Schwarz proprietary and [SCPI conform] value name
RF frequency sweep :TRIGger<hw>:FSWeep:SOURce SINGle [BUS] "Single" or
:TRIGger<hw>:FSWeep:SOURce EXTernal [EXTernal] "Extern Single"
:SOURce<hw>:SWEep:FREQuency:MODE AUTO
RF level sweep :TRIGger<hw>:PSWeep:SOURce SINGle [BUS] for "Single" or
:TRIGger<hw>:PSWeep:SOURce EXTernal [EXTernal] for "Extern Sin-
gle"
:SOURce<hw>:SWEep:POWer:MODE AUTO
LF frequency sweep :TRIGger<hw>:LFFSweep:SOURce SINGle [BUS] "Single" or
:TRIGger<hw>:LFFSweep:SOURce EXTernal [EXTernal] "Extern Sin-
gle"
:SOURce<hw>:LFOutput<ch>:SWEep:FREQuency:MODE AUTO
List :SOURce<hw>:LIST:TRIGger:SOURce SINGle "Single" or
:SOURce<hw>:LIST:TRIGger:SOURce EXTernal "Extern Single"
:SOURce<hw>:LIST:MODE AUTO
Varying the RF Signal in List or Sweep Mode