Configuring the Signal in the RF Domain
R&S
®
SMW200A
463User Manual 1175.6632.02 ─ 16
Step / Extern Step mode (Sweep/List)
Figure 7-12: Step / Extern Step mode (sweep / list )
●
The instrument generates the signal step-by-step.
●
Starts signal generation with "State = On".
●
Switches to the next step with a manual action.
If the end value is reached, signal generation starts again from the beginning.
●
Stops signal generation with "State = Off".
●
Trigger sources:
– The rotary knob at the front panel.
– The ARROW keys at the front panel.
– The corresponding remote control commands.
– An externally applied trigger signal.
To step through the sweep frequencies or levels:
●
In manual mode:
– Set the Current Frequency or Current Level values
– Use the UP and DOWN keys or the rotary knob
●
In remote control mode:
– Use the commands [:SOURce<hw>]:FREQuency:MANual or
[:SOURce<hw>]:POWer:MANual with the UP or DOWN parameter
Steps that would exceed the sweep range are ignored.
Varying the RF Signal in List or Sweep Mode