Remote Control Commands
R&S
®
SMW200A
1043User Manual 1175.6632.02 ─ 16
Parameters:
<Fixed> float
The following settings influence the value range:
An offset set with the command [:SOURce<hw>]:
FREQuency:OFFSet
Numerical value
Sets the frequency in CW and sweep mode
UP|DOWN
Varies the frequency step by step in user mode.
The frequency is increased or decreased by the value set with
the command [:SOURce<hw>]:FREQuency:STEP[:
INCRement].
Range:
(RFmin + OFFSet) to (RFmax + OFFSet)
*RST: 100 MHz
Example: See Example "Frequency configuration" on page 1037
Example: See Example "Setup an RF frequency or power sweep"
on page 1107
Manual operation: See "Frequency" on page 431
[:SOURce<hw>]:FREQuency[:CW|FIXed]:RCL <Rcl>
Set whether the RF frequency value is retained or taken from a loaded instrument con-
figuration, when you recall instrument settings with command *RCL.
The selected mode applies to both RF paths, i.e. a specified suffix is ignored.
Parameters:
<Rcl> INCLude | EXCLude
INCLude
Takes the frequency value of the loaded settings.
EXCLude
Retains the current frequency when an instrument configuration
is loaded.
*RST: INCLude
Example:
SOURce1:FREQuency:CW:RCL INCLude
Manual operation: See "Exclude Frequency" on page 632
[:SOURce<hw>]:FREQuency:MANual <Manual>
Sets the frequency and triggers a sweep step manually if SWEep:MODE MAN.
SOURce Subsystem