Remote Control Commands
R&S
®
SMW200A
1044User Manual 1175.6632.02 ─ 16
Parameters:
<Manual> float
You can select any frequency within the setting range, where:
STARt is set with [:SOURce<hw>]:FREQuency:STARt
STOP is set with [:SOURce<hw>]:FREQuency:STOP
OFFSet is set with [:SOURce<hw>]:FREQuency:OFFSet
Range: (STARt + OFFSet) to (STOP + OFFSet)
Increment: 0.01Hz
*RST: 100 MHz
Default unit: Hz
Example: See Example "Setup an RF frequency or power sweep"
on page 1107
Manual operation: See " Current Frequency" on page 478
[:SOURce<hw>]:FREQuency:MULTiplier <Multiplier>
Sets the multiplication factor N
FREQ:MULT
of a subsequent downstream instrument.
The parameters offset f
FREQ:OFFSer
and multiplier N
FREQ:MULT
affect the frequency value
set with the command FREQ.
The query FREQ? returns the value corresponding to the formula:
f
FREQ
= f
RFout
* N
FREQ:MULT
+ f
FREQ:OFFSer
See Chapter 7.3.4.1, "Displayed RF Frequency and Level Values with Downstream
Instruments", on page 440.
Parameters:
<Multiplier> float
Range: 1 to dynamic
Increment: 0.001
*RST: 1
Example: See Example "Frequency configuration" on page 1037
Manual operation: See "Multiplier" on page 432
[:SOURce<hw>]:FREQuency:OFFSet <Offset>
Sets the frequency offset f
FREQ:OFFSet
of a downstream instrument.
The parameters offset f
FREQ:OFFSer
and multiplier N
FREQ:MULT
affect the frequency value
set with the command FREQ.
The query FREQ? returns the value corresponding to the formula:
f
FREQ
= f
RFout
* N
FREQ:MULT
+ f
FREQ:OFFSer
See Chapter 7.3.4.1, "Displayed RF Frequency and Level Values with Downstream
Instruments", on page 440.
SOURce Subsystem