Remote Control Commands
R&S
®
SMW200A
1045User Manual 1175.6632.02 ─ 16
Note: The offset also affects RF frequency sweep.
Parameters:
<Offset> float
Increment: 0.01
*RST: 0
Example: See Example "Frequency configuration" on page 1037
Manual operation: See "Offset" on page 432
[:SOURce<hw>]:FREQuency:CENTer
<Center>
Sets the center frequency of the sweep.
See Chapter 7.6.2.1, "Correlating Parameters in Sweep Mode", on page 467.
Parameters:
<Center> float
Range: 300 kHz to RFmax
Increment: 0.01 Hz
*RST: 300E6
Default unit: Hz
Example: See Example "Setup an RF frequency or power sweep"
on page 1107
Manual operation: See " Center Frequency " on page 478
[:SOURce<hw>]:FREQuency:SPAN <Span>
Sets the sapn of the frequency sweep range.
See Chapter 7.6.2.1, "Correlating Parameters in Sweep Mode", on page 467.
Parameters:
<Span> float
Full freqeuncy range
Increment: 0.01
*RST: 400E6
Example: See Example "Setup an RF frequency or power sweep"
on page 1107
Manual operation: See "Span" on page 478
[:SOURce<hw>]:FREQuency:STARt <Start>
Sets the start frequency for the RF sweep.
See Chapter 7.6.2.1, "Correlating Parameters in Sweep Mode", on page 467.