Remote Control Commands
R&S
®
SMW200A
1042User Manual 1175.6632.02 ─ 16
Parameters:
<Mode> CW | FIXed | SWEep | LIST
CW|FIXed
Sets the fixed frequency mode. CW and FIXed are synonyms.
The instrument operates at a defined frequency, set with com-
mand [:SOURce<hw>]:FREQuency[:CW|FIXed].
SWEep
Sets sweep mode.
The instrument processes frequency (and level) settings in
defined sweep steps.
Set the range and current frequency with the commands:
[:SOURce<hw>]:FREQuency:STARt on page 1045 and [:
SOURce<hw>]:FREQuency:STOP on page 1046,
[:SOURce<hw>]:FREQuency:CENTer on page 1045,
[:SOURce<hw>]:FREQuency:SPAN on page 1045,
[:SOURce<hw>]:FREQuency:MANual on page 1043
LIST
Sets list mode.
The instrument processes frequency and level settings by
means of values loaded from a list.
To configure list mode settings, use the commands of the Chap-
ter 13.18.11, "SOURce:LIST Subsystem", on page 1073.
*RST:
CW
Example: See Example "Frequency configuration" on page 1037
Example: See Example "Setup an RF frequency or power sweep"
on page 1107
Manual operation: See "State (RF frequency sweep)" on page 476
[:SOURce<hw>]:FREQuency[:CW|FIXed] <Fixed>
Sets the frequency of the RF output signal in the selected path.
The effect depends on the selected mode:
In CW mode (FREQ:MODE CW | FIXed), the instrument operates at a fixed fre-
quency.
In sweep mode (FREQ:MODE SWE), the value applies to the sweep frequency.
The instrument processes the frequency settings in defined sweep steps.
In user mode (FREQ:STEP:MODE USER), you can vary the current frequency
step by step.
SOURce Subsystem