Remote Control Commands
R&S
®
SMW200A
1111User Manual 1175.6632.02 ─ 16
[:SOURce<hw>]:SWEep:POWer:STEP[:LOGarithmic] <Logarithmic>
Sets a logarithmically determined step width for the RF level sweep. The level is
increased by a logarithmically calculated fraction of the current level.
See Chapter 7.6.2.1, "Correlating Parameters in Sweep Mode", on page 467.
Parameters:
<Logarithmic> float
The unit dB is manatory.
Range: 0.01 to 139 dB
Increment: 0.01
*RST: 1
Default unit: dB
Example: See Example "Setup an RF frequency or power sweep"
on page 1107.
Manual operation: See " Step " on page 480
[:SOURce<hw>]:SWEep[:FREQuency]:DWELl <Dwell>
Sets the dwell time for a frequency sweep step.
Parameters:
<Dwell> float
Range: 2E-3 to 100
Increment: 100E-6
*RST: 15E-3
Default unit: s
Example: See Example "Setup an RF frequency or power sweep"
on page 1107.
Manual operation: See " Dwell Time " on page 480
[:SOURce<hw>]:SWEep[:FREQuency]:MODE <Mode>
Sets the cycle mode for the frequency sweep.