Remote Control Commands
R&S
®
SMW200A
1110User Manual 1175.6632.02 ─ 16
Parameters:
<Mode> AUTO | MANual | STEP
AUTO
Each trigger triggers exactly one complete sweep.
MANual
The trigger system is not active. You can trigger every step indi-
vidually with the command [:SOURce<hw>]:POWer:MANual.
The level value increases at each step by the value that you
define with [:SOURce<hw>]:POWer:STEP[:INCRement].
Values directly entered with the command [:SOURce<hw>]:
POWer:MANual are not taken into account.
STEP
Each trigger triggers one sweep step only. The level increases
by the value entered with [:SOURce<hw>]:POWer:STEP[:
INCRement].
*RST:
AUTO
Example: See Example "Setup an RF frequency or power sweep"
on page 1107.
Manual operation: See "Mode" on page 476
[:SOURce<hw>]:SWEep:POWer:POINts <Points>
Sets the number of steps within the RF level sweep range.
See Chapter 7.6.2.1, "Correlating Parameters in Sweep Mode", on page 467.
Parameters:
<Points> integer
Range: 2 to Max
Example: See Example "Setup an RF frequency or power sweep"
on page 1107.
[:SOURce<hw>]:SWEep:POWer:SPACing:MODE?
Queries the level sweep spacing. The sweep spacing for level sweeps is always linear.
Return values:
<Mode> LINear
*RST: LINear
Example:
SWE:POW:SPAC:MODE?
queries the sweep spacing for a level sweep at RF output.
Result: "LIN"
linear spacing
Usage: Query only
SOURce Subsystem