Remote Control Commands
R&S
®
SMW200A
1109User Manual 1175.6632.02 ─ 16
[:SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic.............................................. 1114
[:SOURce<hw>]:SWEep[:FREQuency]:STEP[:LINear].....................................................1114
[:SOURce<hw>]:SWEep:RESet[:ALL]............................................................................ 1115
[:SOURce<hw>]:SWEep:POWer:AMODe <AMode>
Selects the power attenuator mode for the level sweep.
Parameters:
<AMode> NORMal | HPOWer
NORMal
Performs the level settings in the range of the built-in attenuator.
HPOWer
Performs the level settings in the high level range.
*RST: NORMal(HighPower)|AUTO
Example:
SWE:POW:AMOD NORM
Selects the high level ranges for level sweep.
[:SOURce<hw>]:SWEep:POWer:DWELl <Dwell>
Sets the dwell time for a level sweep step.
Parameters:
<Dwell> float
Range: 1E-3 to 100
Increment: 100E-6
*RST: 15E-3
Default unit: s
Example: See Example "Setup an RF frequency or power sweep"
on page 1107.
Manual operation: See " Dwell Time " on page 480
[:SOURce<hw>]:SWEep:POWer:MODE <Mode>
Sets the cycle mode for the level sweep.
SOURce Subsystem