Configuring the Signal in the RF Domain
R&S
®
SMW200A
542User Manual 1175.6632.02 ─ 16
R&S NRP sensors that are connected via LAN or use the USBTMC protocol are not
automatically detected. They are detected by the scan search function.
Access:
Select "RF" > "RF Measurement" > "NRP Sensor Mapping".
The dialog lists all detected R&S NRP sensors for selection and mapping. You can
also browse the network for sensors.
The detected sensors are characterized by the used protocol and the correspond-
ing icon. In the "Mapping" column, you can assign the sensor to one of the availa-
ble sensor channels. The list can contain several entries but the R&S SMW can
only use up to four sensors simultaneously.
The remote commands required to define these settings are described in Chap-
ter 13.16, "SENSe, READ, INITiate and SLISt Subsystems", on page 812.
Settings
Sensor Mapping List .................................................................................................. 542
Scan ...........................................................................................................................543
Sensor Mapping List
Displays a list of all sensor entries with information on the sensor name, the used pro-
tocol, the connector and the assigned mapping.