Configuring the Signal in the RF Domain
R&S
®
SMW200A
555User Manual 1175.6632.02 ─ 16
Note: Always turn the RF power off or disconnected from the sensor before zero-
ing, since the function calibrates the power sensor at zero signal power.
The zeroing process takes a few seconds, depending on the power sensor type.
5. Configure additional parameters for the selected sensor as required.
6. Select "State > On".
7. Close the dialog.
To create user correction data with an R&S NRP and the R&S SMW
We assume, that the power sensor is connected, assigned and ready for operation.
1. Select "RF" > "RF Level" > "User Correction".
2. Select "UCOR Data > User Cor. Data".
3. Create a file: "New > Filename" and confirm with "OK".
4. Select the new file with "Select".
5. Select "Edit User Cor. Data".
6. In the "Edit User Correction Data A ucor", select "Edit > Fill".
The "Fill Table" dialog enables you to fill in the values of the columns automatically.
7. To fill in the frequency column:
Improving Level Performance