Configuring the Signal in the RF Domain
R&S
®
SMW200A
556User Manual 1175.6632.02 ─ 16
a) Select "Select Column To Fill > Frequency / Hz"
b) Select "Range > e.g. 15" to determine the number of values.
c) Select "Start Value > e.g. 1.56 GHz".
d) Select "Increment Value > e.g. 1 MHz", to determine the frequency steps.
e) Select "Fill", to insert the values
In the "Edit User Correction Data A ucor" dialog, the table contains the automati-
cally calculated frequency values.
The function also fills the column of the "Power / dB" values with a predefined
value, since empty cells lead to the data loss of the entire line.
8. Select "Fill with Sensor".
The "Fill User Correction Data With Sensor" dialog provides an overview of the
sensor configuration.
9. Select "Execute".
The R&S SMW successively sets each frequency point, reads the measured power
of the sensor and fills in the value in the correction table.