Monitoring Signal Characteristics
R&S
®
SMW200A
599User Manual 1175.6632.02 ─ 16
The measurement is running automatically with sweep time of 1 s.
The dialog displays the measured values in the units of the RF output signal:
"Level = -30 dBm" and "PEP = -19.06 dBm". These values are identical to the
"Level" and "PEP" values displayed in the status bar.
4. In the block diagram, select "Baseband B > IEEE 802.11" and enable the same
settings as in the first baseband.
5. Select "Baseband B > Signal Routing > Route to Path A and Path B".
The signals of "Baseband A" and "Baseband B" are summed and routed to the RF
A output.
6. In the task bar, select "System Config > Setup > Baseband Powers", enable a sec-
ond analogical measurement with "Source > Baseband B" and select "Start".
Two simultaneous measurements are running.
Querying Information on the Baseband Power Levels