Network Operation and Remote Control
R&S
®
SMW200A
737User Manual 1175.6632.02 ─ 16
For information on how to proceed when network failures occur, see Chapter 15.5,
"Resolving Network Connection Failures", on page 1177.
For further information on the functions to read and write to an open session, as
well as the utility applications the software provides, see the R&S VISA user man-
ual.
Starting a remote control session over LAN with R&S VISA
To set the instrument to remote control, you can use the addressed command >R or
send any command from the controller.
1. Start the R&S VISA Tester. Establish the connection to the R&S SMW.
See "Setting up the controller with R&S VISA" on page 734.
2. In the R&S VISA "Basics" tab, enter an SCPI command, e.g. "*IDN?". Confirm with
"Query".
The instrument is switched to remote control when it receives a command from the
controller.
3. Select "Read" to obtain the instrument response.
Tip: If the "Show Log" checkbox is checked R&S VISA displays each VISA function
call in the log-view on the left. If you check the "Write Log" checkbox the log-view
entry is written to the log file as well. You can operate the log-view in two modes:
the "Live Mode" shows only the most recent messages whereas the "View Mode"
allows you to scroll the history.
4. To set, e.g. the frequency, enter SOUR1:FREQ 4 GHz and select "Write".
To check the performed setting, SOUR1:FREQ? and select "Query".
How to Set Up a Remote Control Connection