EasyManua.ls Logo

Rohde & Schwarz SMW200A - Page 770

Rohde & Schwarz SMW200A
1312 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Network Operation and Remote Control
R&S
®
SMW200A
755User Manual 1175.6632.02 ─ 16
c) Select "Show SCPI List".
The list of all SCPI commands is displayed for example for a final check.
Exporting the SCPI list of the instrument state in one step is a fast and convenient
method. Nevertheless, it usually requires postprocessing on an external PC.
How to check an SCPI list
The easiest way to check a list is to execute it. The generator returns a warning if a
setting could not be performed.
However, we recommend that you check the list and possibly rework. It can be that ...
A parameter has not assigned an SCPI command or an element of the user inter-
face has not an assigned parameter. In these cases, :SYST:INF:SCPI 'SCPI
command not available' is entered in the list instead.
Such entries are also detected during execution. The instrument recognizes these
incomplete commands and displays an error message.
A preset has been executed, but several standards subsequently perform some
internal settings that are also assigned to the list with "SCPI Export".
After a preset still some settings are defined, which are then written to the list gen-
erated with "SCPI Export."
Some suggestions on how you can check and revise a list:
1. Search and remove missing command entries.
2. Remove unnecessary content that has been written after a preset.
How to Record / Create SCPI Lists

Table of Contents

Related product manuals