Remote Control Commands
R&S
®
SMW200A
793User Manual 1175.6632.02 ─ 16
:CALibration<hw>:LEVel:LOCal?
Starts the adjustment of the RF level.
Return values:
<Result> 0 | 1 | OFF | ON
Example:
// Start the adjustment
CALibration:LEVel:LOCal?
// Response: 0
// Level adjustment is performed successfully
Usage: Query only
Manual operation: See "Adjust Level At Current Settings" on page 438
:CALibration<hw>:LEVel:STATe <State>
Activates internal level correction.
Parameters:
<State> OFF | ON | 0 | 1
*RST: ON
Example:
CALibration:LEVel:STATe 1
:CALibration<hw>:LEVel[:MEASure]? [<Force>]
Performs all level adjustments in the selected path.
Query parameters:
<Force> "force"
*RST: force
Return values:
<Measure> 0 | 1
Example:
// Start adjustments for maximum level accuracy
CALibration:LEVel:MEASure?"force"
// 0
// Adjustments are performed successfully
Usage: Query only
:CALibration<hw>:LOSCillator:COUPling:LOCal?
Adjusts the internal LO level at the I/Q modulator automatically, when an external LO
signal is fed.
Return values:
<CouplingLevel> 0 | 1 | OFF | ON
*RST: 1
Example:
CALibration:LOSCillator:COUPling:LOCal 1
CALibration Subsystem