Remote Control Commands
R&S
®
SMW200A
809User Manual 1175.6632.02 ─ 16
:OUTPut:ALL[:STATe] <State>
Activates all RF output signals of the instrument.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: n.a. (factory preset: 0)
Example:
OUTP:ALL OFF
switches off all RF output signals.
Manual operation: See "RF On" on page 429
:OUTPut<hw>[:STATe] <State>
Activates the RF output signal.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
OUTP ON
Activates the RF output.
Manual operation: See "RF On" on page 429
:OUTPut<hw>[:STATe]:PON <Pon>
Defines the state of the RF output signal when the instrument is switched on.
Parameters:
<Pon> OFF | UNCHanged
*RST: n.a. (factory preset: UNCHanged)
Example:
OUTP:PON OFF
The RF output is deactivated when the instrument is switched
on.
Manual operation: See " Power-On State " on page 660
:OUTPut<hw>:AMODe <AMode>
Sets the attenuator mode at the RF output.
Note: The setting [:SOURce<hw>]:POWer:ATTenuation:RFOFf:MODE
FATTenuation has higher priority than the attenuator modes FIXed and MANual.
OUTPut Subsystem