Remote Control Commands
R&S
®
SMW200A
831User Manual 1175.6632.02 ─ 16
The instruments are connected to a LAN and there is no active remote channel
from the controller to the external instruments
External RF instruments are connected to the R&S SMW, for example two
R&S SGS.
To control an external instrument from the R&S SMW remotely
Perform the following general steps:
1. Connect the external RF instruments to the R&S SMW.
See:
Chapter 3.9.5, "How to Cable the Instruments in MIMO Test Setups",
on page 197
"To connect instruments to the I/Q analog interface" on page 199
2. Connect a controller and establish a remote control connection to the R&S SMW,
see Chapter 12.8, "How to Set Up a Remote Control Connection", on page 732.
3. Write an application program and use the
SCONfiguration:EXTernal:<Connector>:REMote:SEND commands to con-
trol the connected RF instruments.
See also Example "Connecting an external instrument and sending SCPI com-
mands to it" on page 832.
SCONfiguration Subsystem