Remote Control Commands
R&S
®
SMW200A
839User Manual 1175.6632.02 ─ 16
SCONfiguration:MIMO:SUBSet SET2
// trigger the baseband signal generation and hence the fading process
OUTPut1:USER6:TRIGger:IMMediate
Options: R&S SMW-K821
Manual operation: See "Subset" on page 142
:SCONfiguration:BASeband:SOURce <SourConfig>
Determines whether coupled or separated baseband sources are used.
Parameters:
<SourConfig> SEParate | COUPled | CPENtity
SEParate
Enabled in LxMxN configurations with L ≤ 2 and M < 4
COUPled|CPENtity
Enabled in LxMxN configurations with L > 2
CPENtity enabled in LxMxN configurations with L > 1 and M >
1.
*RST: SEParate
Example: See Example "Distributing the I/Q streams to the output connec-
tors (advanced configuration)" on page 826
Manual operation: See "Baseband Source Configuration" on page 143
:SCONfiguration:DUPLicate[:STReam] <DuplicateStream>
In a 3x1x1 or 4x1x1 configuration, creates a copy of each stream. Generates are a
total number of 6 or 8 streams, where 4 of them can be signals with real-time data
source.
Parameters:
<DuplicateStream> 0 | 1 | OFF | ON
*RST: 0
Example:
:SCONfiguration:APPLy
:SCONfiguration:MODE ADV
:SCONfiguration:FADing SISO3X1X1
:SCONfiguration:DUPLicate:STReam 1
:SCONfiguration:APPLy
Options: R&S SMW-K550
Manual operation: See "Duplicate Streams" on page 143
:SCONfiguration:APPLy
Assigns and confirms the settings.
SCONfiguration Subsystem