Remote Control Commands
R&S
®
SMW200A
855User Manual 1175.6632.02 ─ 16
:SCONfiguration:EXTernal:BBMM<ch>:REMote:SEND <SendScpiCommand>
:SCONfiguration:EXTernal:IQOutput<ch>:REMote:SEND <SendScpiCommand>
Allows you to send SCPI commands to the RF instruments connected to the
R&S SMW.
Setting parameters:
<SendScpiCommand>"<SCPI syntax>"
String containing an SCPI command (query or setting)
Example: See Chapter 13.17, "SCONfiguration Subsystem", on page 825
Usage: Setting only
Manual operation: See "Command, Send Command" on page 162
:SCONfiguration:EXTernal:IQOutput<ch>:CONNections:CHECk?
Queries the status of the required connections between the R&S SMW and the R&S
SZU.
R&S SZU is connected to the R&S SMW via USB.
Return values:
<ConnectionStati> <RF_conn_status>,<I_OUT_conn_sta-
tus>,<I_BAR_OUT_conn_status>,<Q_OUT_conn_sta-
tus>,<Q_BAR_OUT_conn_status>,<Trigger_conn_status>
Unused
Connection not used and not required.
Failed
The physical connection is required but missing or is incorrect.
Passed
The required physical connection is up and running.
Example:
SCONfiguration:EXTernal:IQOutput1:REMote:ISELect "SZU (999991) B1066"
SOURce1:IQ:OUTPut:ANALog:TYPE SING
SCONfiguration:EXTernal:IQOutput1:REMote:CONNect
SCONfiguration:EXTernal:ACONnect 1
SCONfiguration:EXTernal:PBEHaviour 1
SCONfiguration:EXTernal:IQOutput1:REMote:INFO?
// "SZU (999991) B1066","USB (vendor specific)","999991","A"
SOURce1:FREQuency:CONVerter:EXTernal:CHECk:CONNections?
// Passed,Failed,Unused,Failed,Unused,Passed
// The R&S SZU is connected to the R&S SMW
// The USB and the RF A to LO In connections are correct
// There is a failure in the single-ended I/Q connection
// (between the I/Q OUT and the I/Q IN connectors)
Usage:
Query only
Manual operation:"Check Connections Settings" on page 163
SCONfiguration Subsystem