EasyManua.ls Logo

Rohde & Schwarz SMW200A - Page 894

Rohde & Schwarz SMW200A
1312 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMW200A
879User Manual 1175.6632.02 ─ 16
Parameters:
<Mode> HBANdwidth | HDEViation | LNOise
HBANdwidth
Sets the maximum available bandwidth.
HDEViation
Sets the maximum range for ΦM deviation.
LNOise
Selects a phase modulation mode with phase noise and spuri-
ous characteristics close to CW mode.
*RST: HBANdwidth
Example: See Example "Performing phase modulation" on page 877.
Manual operation: See "Mode" on page 510
[:SOURce<hw>]:PM:RATio
<Ratio>
Sets the deviation ratio (path2 to path1) in percent.
Parameters:
<Ratio> float
Range: 0 to 100
Increment: 0.01
*RST: 100
Example: See Example "Performing phase modulation" on page 877.
Manual operation: See "Ratio Path2/Path1" on page 509
[:SOURce<hw>]:PM:SENSitivity?
Queries the sensitivity of the externally applied signal for phase modulation.
The returned value reports the sensitivity in RAD/V. It is assigned to the voltage value
for full modulation of the input.
Return values:
<Sensitivity> float
Example: See Example "Performing phase modulation" on page 877.
Usage: Query only
[:SOURce<hw>]:PM<ch>:SOURce <Source>
Selects the modulation source for phase modulation signal.
Suffix:
PM<ch>
.
1|2
Sets the modulation signal channel.
SOURce Subsystem

Table of Contents