Remote Control Commands
R&S
®
SMW200A
880User Manual 1175.6632.02 ─ 16
Parameters:
<Source> LF1 | LF2 | NOISe | EXT1 | EXT2 | INTernal | EXTernal
LF1|LF2
Uses an internally generated LF signal.
EXT1|EXT2
Uses an externally supplied LF signal.
NOISe
Uses the internally generated noise signal.
INTernal
Uses the internally generated signal of LF1.
EXTernal
Uses an external LF signal (EXT1).
*RST:
LF1 <PM1>; LF2 <PM2>
Example: See Example "Performing phase modulation" on page 877.
Manual operation: See "Source" on page 507
[:SOURce<hw>]:PM<ch>:STATe <State>
Activates phase modulation.
Activation of phase modulation deactivates frequency modulation.
Suffix:
PM<ch>
.
1|2
Sets the modulation signal channel.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Performing phase modulation" on page 877.
Manual operation: See "State" on page 506
[:SOURce]:PM<ch>[:DEViation] <Deviation>
Sets the modulation deviation of the phase modulation in RAD.
Parameters:
<Deviation> float
The maximal deviation depends on the RF frequency and the
selected modulation mode (see data sheet).
Range: 0 to max
Increment: 1
*RST: 1
Default unit: RAD
Example: See Example "Performing phase modulation" on page 877.
Manual operation: See "Deviation" on page 509