Configuring the Signal in the RF Domain
R&S
®
SMW200A
507User Manual 1175.6632.02 ─ 16
Remote command:
[:SOURce<hw>]:AM<ch>:STATe on page 873
[:SOURce<hw>]:FM<ch>:STATe on page 877
[:SOURce<hw>]:PM<ch>:STATe on page 880
Source
Selects the LF signal source.
You can vary the signal sources for each of the modulations.
"LF Generator1/2"
Selects one of internally generated LF signals.
"External 1/2"
Selects an externally supplied LF signal.
"Noise Generator"
Selects the internally generated noise signal.
Remote command:
[:SOURce<hw>]:AM<ch>:SOURce on page 872
[:SOURce<hw>]:FM<ch>:SOURce on page 876
[:SOURce<hw>]:PM<ch>:SOURce on page 879
AM
Provides the settings for amplitude modulation.
AM Depth ← AM
Determines the depth of the modulation signal in percent.
The depth is limited by the maximum peak envelope power (PEP).
Remote command:
[:SOURce<hw>]:AM<ch>[:DEPTh] on page 873
Ratio Path2/Path1 ← AM
Sets the deviation ratio (path2 to path1) in per cent.
Remote command:
[:SOURce<hw>]:AM:RATio on page 872