Remote Control Commands
R&S
®
SMW200A
889User Manual 1175.6632.02 ─ 16
Parameters:
<Mode> CODer | BBIN
CODer
Internal baseband generator
BBIN
An external digital baseband signal provided at the digital inter-
face
*RST: CODer
Example:
SOURce1:BB:CODer:MODE BBIN
[:SOURce<hw>]:BB:POWer:PEAK?
Queries the peak level of the baseband signal relative to full scale of 0.5 V (in terms of
dB full scale).
Return values:
<Peak> float
Range: -145 to 30
Increment: 0.01
*RST: 0
Default unit: dBfs
Example:
BB:POW:PEAK
Queries the peak level of the baseband signal.
Usage: Query only
[:SOURce<hw>]:BB:CFACtor?
Queries the crest factor of the baseband signal.
Return values:
<CFactor> float
Range: 0 to 100
Increment: 0.01
*RST: 0
Default unit: dB
Example:
SOURce1:BB:CFACtor?
queries the crest factor of the baseband signal.
Usage: Query only
[:SOURce<hw>]:BB:POWer:RMS?
Queries the RMS level of the baseband signal relative to full scale of 0.5V (in terms of
dB full scale).
SOURce Subsystem