Remote Control Commands
R&S
®
SMW200A
951User Manual 1175.6632.02 ─ 16
Example:
BB:ARB:MCAR:EDIT:CARR:STAT ON
Sets all the carriers in the carrier range to ON.
Manual operation: See "Carrier State" on page 368
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:STARt <Start>
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:STOP <Stop>
Selects the last carrier in the carrier range to which the settings shall apply.
Parameters:
<Stop> integer
Range: 0 to 511
*RST: 0
Example:
BB:ARB:MCAR:EDIT:CARR:STOP 4
The carrier range stops at carrier 4.
Manual operation: See "Carrier Start/Start" on page 368
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:POWer:STEP <Step>
Sets the step width by which the starting power of the carriers in the defined carrier
range is incremented.
Parameters:
<Step> float
Range: -80 to 80
Increment: 0.01
*RST: 0
Default unit: dB
Example:
BB:ARB:MCAR:EDIT:CARR:POW -80dB
Sets a power of -80 dB for the carriers in the carrier range.
BB:ARB:MCAR:EDIT:CARR:POW:STEP 1 dB
The power is incremented by 1dB for each carrier.
That is, the first carrier has -80 dB, the second -79 dB, and so
on.
Manual operation: See "Gain Step" on page 368
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:POWer[:STARt] <Start>
Sets the power for the individual carriers in the defined carrier range.
Parameters:
<Start> float
Range: -80 to 0
Increment: 0.01
*RST: 0
Default unit: dB
SOURce Subsystem