Remote Control Commands
R&S
®
SMW200A
952User Manual 1175.6632.02 ─ 16
Example:
BB:ARB:MCAR:EDIT:CARR:POW -50 dB
sets the power of the carriers in the carrier range to -50 dB.
Manual operation: See "Gain Start" on page 368
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:PHASe:STEP <Step>
Sets the step width by which the start phases of the carriers in the defined carrier
range is incremented.
Parameters:
<Step> float
Range: -359.99 to 359.99
Increment: 0.01
*RST: 0
Default unit: DEG
Example:
BB:ARB:MCAR:EDIT:CARR:PHAS 90 DEG
Sets a start phase of 90° for the carriers in the carrier range.
BB:ARB:MCAR:EDIT:CARR:PHAS:STEP 1 DEG
The start phase is incremented by 1° for each carrier. That is,
the first carrier has a start phase of 90°, the second a start
phase of 91°, and so on.
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:PHASe[:STARt] <Start>
Sets the start phase for the individual carriers in the defined carrier range.
Parameters:
<Start> float
Range: 0 to 359.99
Increment: 0.01
*RST: 0
Default unit: DEG
Example:
BB:ARB:MCAR:EDIT:CARR:PHAS 90 DEG
sets a start phase of 90° for the carriers in the carrier range.
Manual operation: See "Phase Start" on page 368
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:DELay:STEP <Step>
Sets the step width by which the start delays of the carriers in the defined carrier range
is incremented.
Parameters:
<Step> float
Range: -1 to 1
Increment: 1E-9
*RST: 0
Default unit: s
SOURce Subsystem