Remote Control Commands
R&S
®
SMW200A
969User Manual 1175.6632.02 ─ 16
Setting parameters:
[TRACE] MARKER | BURST | LEVATT | CW MODE | HOP | MAP
Name of the marker or control signal.
For ARB waveforms, it is only meaningful to define marker sig-
nals; in the ARB multi-segment waveforms these tags are
ignored!
[#] 1 .. 3
Sets the marker or control trace number; supported is only
LEVATT LIST 1.
Pos Specifies in ASCII format the position (i.e. sample number or
data value), with effect from which the binary State of the
marker or of the control signal changes.
State 0 | 1
Specifies the binary state of the marker or of the control signal
from Pos
x
to Pos
x+1
exclusive in ASCII format.
Example:
{MARKER LIST 1: 0:0;10:1;20:0;30:1}
BB:DM:CLIS:TAG? 'MARKER LIST 1'
Queries the content of the MARKER LIST 1 tag of the selected
control list file.
Response: '0:0;10:1;20:0;30:1'
The marker setting for samples 0 to 9 = 0 (low), for 10 to 19 = 1
(high) and for 20 to 29 = 0. From sample 30 onward the marker
setting = 1.
Example:
{LEVATT LIST 1: 0:0;10:1;20:0;30:1}
BB:DM:CLIS:TAG? 'LEVATT LIST 1'
Queries the content of the LEVATT LIST 1 tag of the selected
control list file.
Response: '0:0;10:1;20:0;30:1'
Level attenuation applies to data values 10 to 19 (high) and from
data value 30 onward.
Usage: Setting only
Manual operation: See "Select Ramp to Edit" on page 283
{WAVEFORM-Length: #I0Q0I1Q1...IxQx...IN-1QN-1…}
(mandatory for waveforms)
The tag contains the actual waveform data or multi-segment waveform data (I/Q
stream). Refer to Chapter 4.8, "Generating Multi-Segment Waveform Files",
on page 324 for background information description of the multi-segment waveform
function.