Remote Control Commands
R&S
®
SMW200A
1020User Manual 1175.6632.02 ─ 16
// Response: 0
// performed is one single measurment
// query baseband power values
:SOURce:BB:MEASurement:POWer:PEAK?
// Response: -24.0601192007,-24.0601192007,-24.0601192007,-24.0601192007
:SOURce:BB:MEASurement:POWer:RMS?
// Response: -35.0006514336062,-35.0006514336062,-35.0006514336062,-35.0006514336062
// performed are four sub-measurements
// stop and remove the measurement
:SOURce:BB:MEASurement:POWer:INDex 1
:SOURce:BB:MEASurement:POWer:ABORt
:SOURce:BB:MEASurement:POWer:DELete
[:SOURce]:BB:MEASurement:POWer:SOURce <Source>
Defines the measurement signal source.
Parameters:
<Source> BBA | BBB | BBC | BBD | BBINA | BBINB | FADINPA |
FADINPB | FADINPC | FADINPD | FADOUTA | FADOUTB |
FADOUTC | FADOUTD | AWGNA | AWGNB | AWGNC |
AWGND | STREAMA | STREAMB | STREAMC | STREAMD
*RST: BBA
Example: See Example "Querying information on burst baseband signals
using the multi-gated acquisition" on page 1019
Manual operation: See "Source/Output" on page 595
[:SOURce]:BB:MEASurement:POWer:OUTPut <Output>
Defines the output point the measurement are performed at.
Parameters:
<Output> RFA | RFB | IQOUT1 | IQOUT2 | BBMM1 | BBMM2
*RST: RFA
Example: See Example "Querying information on burst baseband signals
using the multi-gated acquisition" on page 1019
Manual operation: See "Source/Output" on page 595
[:SOURce]:BB:MEASurement:POWer:ACQuisition <Acquistion>
Sets the acquisition method.
Parameters:
<Acquistion> NOMinal | CONTinuous | GATed | MGATed
*RST: CONTinuous
SOURce Subsystem