Remote Control Commands
R&S
®
SMW200A
1079User Manual 1175.6632.02 ─ 16
Parameters:
<Dwell> <Dwell#1>{, <Dwell#2>, ...} | block data
You can either enter the data as a list of numbers, or as binary
block data. The list of numbers can be of any length, with the list
entries separated by commas.
In binary block format, 8 (4) bytes are always interpreted as a
floating-point number with double accuracy. See also :
FORMat[:DATA] on page 802 for more details.
Example: See Example "Create an RF list and activate the list mode"
on page 1074.
Manual operation: See "Edit List Mode Data" on page 487
[:SOURce<hw>]:LIST:DWELl:LIST:POINts?
Queries the number (points) of dwell time entries in the selected list.
Return values:
<Points> integer
Range: 0 to INT_MAX
*RST: 0
Example: See Example "Create an RF list and activate the list mode"
on page 1074.
Usage: Query only
[:SOURce<hw>]:LIST:FREQuency <Frequency>
Enters the frequency values in the selected list.
Parameters:
<Frequency> <Frequency#1>{, <Frequency#2>, ...} | block data
You can either enter the data as a list of numbers, or as binary
block data.
The list of numbers can be of any length, with the list entries
separated by commas.
In binary block format, 8 (4) bytes are always interpreted as a
floating-point number with double accuracy.
See also :FORMat[:DATA].
Range: 300 kHz to RFmax (depends on the installed
options)
Example: See Example "Create an RF list and activate the list mode"
on page 1074.
Manual operation: See "Edit List Mode Data" on page 487
SOURce Subsystem