Remote Control Commands
R&S
®
SMW200A
1084User Manual 1175.6632.02 ─ 16
[:SOURce<hw>]:LIST:CATalog?
Queries the available list files in the specified directory.
Return values:
<Catalog> string
List of list filenames, separated by commas
Example: See Example "Create an RF list and activate the list mode"
on page 1074.
Usage: Query only
Manual operation: See " List Mode Data " on page 484
[:SOURce<hw>]:LIST:DELete
<Filename>
Deletes the specified list.
Setting parameters:
<Filename> string
Filename or complete file path; file extension is optional.
Example: See [:SOURce<hw>]:LIST:DELete:ALL on page 1084.
Usage: Setting only
Manual operation: See " List Mode Data " on page 484
[:SOURce<hw>]:LIST:DELete:ALL
Deletes all lists in the set directory.
This command can only be executed, if:
No list file is selected.
List mode is disabled.
Example:
SOUR1:LIST:CAT?
// list,my_list
SOUR1:LIST:DEL "/var/user/list1"
SOUR1:LIST:CAT?
// my_list
SOUR1:FREQ:MODE?
// LIST
SOUR1:LIST:SEL?
// /var/user/my_list.lsw
//deactivate list mode
SOUR1:FREQ:MODE CW
SOUR1:LIST:DELete:ALL
SOUR1:LIST:CAT?
// -
// all list files are deleted
SOURce Subsystem