Remote Control Commands
R&S
®
SMW200A
1145User Manual 1175.6632.02 ─ 16
Parameters:
<IqSource> SINE | CONStant | ARB
SINE
Generates a sine waveform. To determine the frequency, use
command :TEST:BB:GENerator:FREQuency on page 1144.
CONStant
Uses a constant I/Q test signal.
ARB
Selects a signal generated generated by the ARB.
*RST: SINE
Example:
:TEST:BB:GENerator:SOURce SINE
Manual operation: See "IQ Source" on page 1165
:TEST:BB:GENerator:STATe <State>
Starts the test generator.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
TEST:BB:GENerator:STATe ON
// activates the test generator
Manual operation: See "State" on page 1165
:TEST<hw>:DIRect:BLOCk:READ? <Board>, <SubAddr>, <Addr>[, <Len>]
Direct mode for querying the blocks of the hardware assemblies directly.
The respective hardware assembly responds directly to the command; any safety
mechanisms are bypassed. This function is only available via remote control.
Query parameters:
<Board> string
<SubAddr> integer
<Addr> integer
<Len> integer
Return values:
<DataBlock> string
Example:
TEST:DIR:BLOC:READ? 'TEST_SCPI',0,0,8
Usage: Query only
TEST Subsystem