General Instrument Functions
R&S
®
SMW200A
671User Manual 1175.6632.02 ─ 16
2. Use the built-in Show Connector function to display the physical location of the
selected connector.
A blinking LED on the rear panel indicates the selected connector, too.
Show Connector
Accesses a dialog that displays the physical location of the selected connector on the
front/rear panel of the instrument.
A blinking LED on the front/rear panel indicates the selected connector, too (see "Con-
nector status indication" on page 663).
User 1 .. 6 Connector Direction
Determines whether the connector is used as an input or an output.
A connector with not specified direction is in a "Not Used" state.
Remote command:
[:SOURce]:INPut:USER<ch>:DIRection on page 858
:OUTPut:USER<ch>:DIRection on page 858
Signal
Determines the control signal that is input or output at the selected connector. Whether
the connector is used as an input or an output is selected by the parameter Direction
The Table 11-1 lists the available control signals.
"Global Trigger 1/2"
Input for triggering custom digital modulations, digital standards and
ARB. A trigger event affects the signal generation in all basebands for
which the "Global Trigger" signal is enabled.