Remote Control Commands
R&S
®
SMW200A
859User Manual 1175.6632.02 ─ 16
[:SOURce<hw>]:INPut:TM<ch>:SIGNal <Signal>
Determines the control signal that is input at the selected connector.
To define the connector direction, use the command [:SOURce<hw>]:INPut:
TM<ch>:DIRection.
Suffix:
SOURce<hw>
.
1 | 2
Determines the baseband the connectors belong to, where:
SOURce1 = Baseband A/C and
SOURce2 = Baseband B/D
TM<ch> 1|2|3
Determines the local connector, where:
SOURce1:INPut:TM1|2|3 = TMC1/2/3 of Baseband A/C and
SOURce2:INPut:TM1|2|3 = TMC4/5/6 of Baseband B/D
Parameters:
<Signal> TRIGger | CLOCk | FEEDback | DATA
CLOCk is available only for TM1
DATA is available only for TM2
FEEDback is available only for TM3
Manual operation: See "Signal" on page 675
:OUTPut<hw>:TM<ch>:SIGNal <Signal>
Determines the control signal that is output at the selected connector.
To define the connector direction, use the command :OUTPut<hw>:TM<ch>:
DIRection.
Suffix:
OUTPut<hw>
.
1 | 2
Determines the baseband the connectors belong to, where:
OUTPut1 = Baseband A/C or Baseband 1/3
OUTPut2 = Baseband B/D or Baseband 2/4
TM<ch> 1|2|3
Determines the local connector, where:
OUTPut1:TM1|2|3 = TMC1/2/3 of Baseband A/C or 1/3 and
OUTPut2:TM1|2|3 = TMC4/5/6 of Baseband B/D or 2/4
SOURce Subsystem