Remote Control Commands
R&S
®
SMW200A
860User Manual 1175.6632.02 ─ 16
Parameters:
<Signal> MARKA1 | MARKA2 | MARKA3 | SCLA | LATTA | BGATA |
HOPA | CWMODA | TRIGA | MARKB1 | MARKB2 | MARKB3 |
SCLB | LATTB | BGATB | HOPB | CWMODB | TRIGB |
MARKC1 | MARKC2 | MARKC3 | SCLC | LATTC | BGATC |
HOPC | CWMODC | TRIGC | MARKD1 | MARKD2 | MARKD3 |
SCLD | LATTD | BGATD | HOPD | CWMODD | TRIGD
MARKA1|MARKC1|MARKA2|MARKC2|MARKA3|MARKC3 = Base-
band A/C Marker 1/2/3
SCLA|SCLB|SCLC|SCLD = Symbol Clock A/B/C/D
LATTA|LATTB|LATTC|LATTD = Lev Att A/B/C/D
BGATA|BGATB|BGATC|BGATD = Burst Gate A/B/C/D
HOPA|HOPB|HOPC|HOPD = HOP A/B/C/D
CWMODA|CWMODB|CWMODC|CWMODD = CW/Mod A/B/C/D
TRIGA|TRIGB|TRIGC|TRIGD = Triggered A
The character A/B/C/D in the parameter value indicates the
baseband the signal is related to.
*RST:
depends on TM suffix
Example:
:SOURce2:INPut:TM2:DIRection OUTPut
:OUTPut2:TM2:DIRection?
Response: OUTPut
:OUTPut2:TM2:SIGNal MARKA2
Manual operation: See "Signal" on page 675
[:SOURce]:INPut:USER<ch>:SIGNal <Signal>
Determines the control signal that is input at the selected connector.
To define the connector direction, use the command [:SOURce]:INPut:USER<ch>:
DIRection.
Suffix:
USER<ch>
.
1 to 6
Parameters:
<Signal> TRIG1 | TRIG2 | NSEGM1 | NSEGM2 | IPULSA | IPULSB |
NONE | FEEDback | SYNCIN
TRIG1|TRIG2 = Global Trigger 1/2
NSEGM1|NSEGM2 = Global Next Segment 1/2
IPULSA|IPULSB = Pulse In A/B, available for USER4|5|6
FEEDback = Baseband Feedback, available for USER6
SYNCIN = Baseband Sync In
Example: See Example "Global connectors settings configuration"
on page 857
Options: NSEGM1|NSEGM2 require R&S SMW-B10
SYNCIN requires R&S SMW-B9
Manual operation: See "Signal" on page 671
SOURce Subsystem