Remote Control Commands
R&S
®
SMW200A
884User Manual 1175.6632.02 ─ 16
[:SOURce<hw>]:PULM:IMPedance <Impedance>
Sets the impedance for the external pulse modulation input.
Parameters:
<Impedance> G50 | G1K
*RST: G1K
Example: See Example "Perform pulse modulation" on page 881.
Manual operation: See "Impedance Pulse Input" on page 514
[:SOURce<hw>]:PULM:MODE
<Mode>
Selects the mode for the pulse modulation.
Parameters:
<Mode> SINGle | DOUBle
SINGle
Generates a single pulse.
DOUBle
Generates two pulses within one pulse period.
*RST: SINGle
Example: See Example "Perform pulse modulation" on page 881.
Manual operation: See "Pulse Mode" on page 515
[:SOURce<hw>]:PULM:OUTPut:VIDeo:POLarity <Polarity>
Sets the polarity of the pulse video (modulating) signal, related to the RF (modulated)
signal.
Parameters:
<Polarity> NORMal | INVerted
NORMal
the video signal follows the RF signal, that means it is high
wihen RF signal is high and vice versa.
INVerted
the video signal follows in inverted mode.
*RST: NORMal
Example: See Example "Perform pulse modulation" on page 881.
Manual operation: See "Video Polarity" on page 514
SOURce Subsystem