Getting Started
R&S
®
SMW200A
75User Manual 1175.6632.02 ─ 16
To reconfigure the trigger settings
We assume that the instrument is configured as described in Chapter 2.3.2, "Generat-
ing a Digitally Modulated Signal", on page 69 and the default connector mapping is
maintained (see Figure 2-9).
1. In the block diagram, select "Baseband > Misc > Custom Digital Mod > Trigger In".
2. Select the following settings:
a) "Mode > Armed Auto"
b) "Source > External Global Trigger 1".
The instrument expects an external global trigger event. In this current configura-
tion (see Figure 2-9), the "Global Trigger 1" signal has to be supplied at the input
connector USER 3.
The Trigger/Marker/Clock status LEDs in the block diagram confirm that an exter-
nal trigger signal is selected; the signal generation is however stopped.
Trying Out the Instrument