Remote Control Commands
R&S
®
SMW200A
955User Manual 1175.6632.02 ─ 16
Return values:
<RMode> STOP | RUN
Example: see Example "Managing waveform files" on page 920
Usage: Query only
Manual operation: See "Running/Stopped" on page 255
[:SOURce<hw>]:BB:ARBitrary:TRIGger:SLENgth <SLength>
Defines the length of the signal sequence that is output in the SINGle trigger mode.
Parameters:
<SLength> integer
Maximun value dependents on the selected units [:
SOURce<hw>]:BB:ARBitrary:TRIGger:SLUNit as follows:
SAMPle: Max = 2
32
-1
SEQuence: Max = 1000
Range: 1 to dynamic
*RST: 1
Example: See Example "Managing waveform files" on page 920
[:SOURce<hw>]:BB:ARBitrary:TRIGger:SLUNit <SLUnit>
Defines the unit for the entry of the length of the signal sequence to be output in the
Single trigger mode.
Parameters:
<SLUnit> SEQuence | SAMPle
*RST: SEQuence
Example: See Example "Managing waveform files" on page 920
Manual operation: See "Trigger Signal Duration" on page 255
[:SOURce<hw>]:BB:ARBitrary:TRIGger:ARM:EXECute
Stops (arms) waveform output.
Example:
See Example "Managing waveform files" on page 920
Usage: Event
Manual operation: See "Arm" on page 255
[:SOURce<hw>]:BB:ARBitrary:TRIGger:EXECute
Executes an internal trigger event.
Example:
See Example "Managing waveform files" on page 920
SOURce Subsystem