Remote Control Commands
R&S
®
SMW200A
998User Manual 1175.6632.02 ─ 16
Provided only for backward compatibility with other Rohde & Schwarz signal gener-
ators.
The R&S SMW accepts these values and maps them automatically as follows:
EXTernal = EGT1, BEXTernal = EGT2, OBASeband = INTA or INTB
(depending on the current baseband)
Parameters:
<Source> INTB | INTernal | OBASeband | EGT1 | EGT2 | EGC1 | EGC2 |
ELTRigger | INTA | ELCLock | BEXTernal | EXTernal | BBSY
*RST: INTernal
Example: See "Programming Examples" on page 983
Options: ELTRigger|ELCLock require R&S SMW-B10
BBSY require R&S SMW-B9
Manual operation: See "Trigger Source" on page 256
[:SOURce<hw>]:BB:MCCW:TRIGger[:EXTernal]:DELay <Delay>
Specifies the trigger delay.
Parameters:
<Delay> float
Range: 0 to 2147483647
Increment: 0.01
*RST: 0
Default unit: samples
Example: See "Programming Examples" on page 983
Manual operation: See "(Specified) External Delay/(Specified) Trigger Delay"
on page 257
[:SOURce<hw>]:BB:MCCW:TRIGger[:EXTernal]:INHibit <Inhibit>
Specifies the number of samples by which a restart is inhibited following an external
trigger event.
Parameters:
<Inhibit> integer
Range: 0 to 21.47*clockRate
*RST: 0
Example: See "Programming Examples" on page 983
Manual operation: See "External / Trigger Inhibit" on page 257
[:SOURce<hw>]:BB:MCCW[:TRIGger]:SEQuence <Sequence>
Selects the trigger mode. For detailed description of the trigger modes, refer to "Impact
of the Trigger Modes on the Signal Generation" on page 239.
SOURce Subsystem