Remote Control Commands
R&S
®
SMW200A
1113User Manual 1175.6632.02 ─ 16
Parameters:
<Spacing> LINear | LOGarithmic
LINear
Sets a fixed frequency value as step width and adds it to the cur-
rent frequency.
The linear step width is entered in Hz, see [:SOURce<hw>]:
SWEep[:FREQuency]:STEP[:LINear].
LOGarithmic
Sets a constant fraction of the current frequency as step width
and adds it to the current frequency.
The logarithmic step width is entered in %, see [:
SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic.
*RST:
LINear
Example: See Example "Setup an RF frequency or power sweep"
on page 1107.
Manual operation: See " Spacing" on page 479
[:SOURce<hw>]:SWEep:POWer:SHAPe <Shape>
[:SOURce<hw>]:SWEep[:FREQuency]:SHAPe <Shape>
Determines the waveform shape for a frequency sweep sequence.
Parameters:
<Shape> SAWTooth | TRIangle
*RST: SAWTooth
Example: See Example "Setup an RF frequency or power sweep"
on page 1107.
Manual operation: See " Shape " on page 479
[:SOURce<hw>]:SWEep:POWer:EXECute
[:SOURce<hw>]:SWEep[:FREQuency]:EXECute
Executes an RF frequency sweep.
The command performs a single sweep and is therefore only effective in manual
sweep mode.
Example:
See Example "Setup an RF frequency or power sweep"
on page 1107.
Usage: Event
Manual operation: See " Execute Single Sweep " on page 477
SOURce Subsystem

Table of Contents