EasyManua.ls Logo

Rohde & Schwarz SMW200A - Page 1158

Rohde & Schwarz SMW200A
1312 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMW200A
1143User Manual 1175.6632.02 ─ 16
Parameters:
<BncDestination> AUTO | USER1 | USER2 | USER3 | USER4 | USER5 | USER6 |
TRGA | TRGB | C1TMC1 | C1TM2 | C1TM3 | C2TMC4 |
C2TM5 | C2TM6 | F1TMC1 | F1TM2 | F1TM3 | F2TMC4 |
F2TM5 | F2TM6 | F3TMC1 | F3TM2 | F3TM3 | F4TMC4 |
F4TM5 | F4TM6
*RST: USER2
Example:
SYSTem:PROTect1 0,123456
TEST:BB:BNC:DESTination TRGA
// tests the instrument trigger connector of path A
Manual operation: See "Destination" on page 1164
:TEST:BB:BNC:LOG?
Queries the log message reported during the BNC connector test.
This is a password-protected function. Unlock the protection level 1 to access it, see :
SYSTem:PROTect<ch>[:STATe].
Return values:
<Log> string
Example:
SYSTem:PROTect1 0,123456
TEST:BB:BNC:LOG?
// queries the reported message
Usage: Query only
Manual operation: See "Log Message" on page 1164
:TEST:BB:BNC:SOURce <BncSource>
Selects the BNC connection test source.
This is a password-protected function. Unlock the protection level 1 to access it, see :
SYSTem:PROTect<ch>[:STATe].
Parameters:
<BncSource> AUTO | USER1 | USER2 | USER3 | USER4 | USER5 | USER6 |
TRGA | TRGB | C1TMC1 | C1TM2 | C1TM3 | C2TMC4 |
C2TM5 | C2TM6 | F1TMC1 | F1TM2 | F1TM3 | F2TMC4 |
F2TM5 | F2TM6 | F3TMC1 | F3TM2 | F3TM3 | F4TMC4 |
F4TM5 | F4TM6
*RST: USER1
Example:
SYSTem:PROTect1 0,123456
TEST:BB:BNC:SOURce AUTO
// detects the connectors that have assigned a signal and performs the connectror test
Manual operation: See "Source" on page 1164
TEST Subsystem

Table of Contents

Related product manuals