Remote Control Commands
R&S
®
SMW200A
1143User Manual 1175.6632.02 ─ 16
Parameters:
<BncDestination> AUTO | USER1 | USER2 | USER3 | USER4 | USER5 | USER6 |
TRGA | TRGB | C1TMC1 | C1TM2 | C1TM3 | C2TMC4 |
C2TM5 | C2TM6 | F1TMC1 | F1TM2 | F1TM3 | F2TMC4 |
F2TM5 | F2TM6 | F3TMC1 | F3TM2 | F3TM3 | F4TMC4 |
F4TM5 | F4TM6
*RST: USER2
Example:
SYSTem:PROTect1 0,123456
TEST:BB:BNC:DESTination TRGA
// tests the instrument trigger connector of path A
Manual operation: See "Destination" on page 1164
:TEST:BB:BNC:LOG?
Queries the log message reported during the BNC connector test.
This is a password-protected function. Unlock the protection level 1 to access it, see :
SYSTem:PROTect<ch>[:STATe].
Return values:
<Log> string
Example:
SYSTem:PROTect1 0,123456
TEST:BB:BNC:LOG?
// queries the reported message
Usage: Query only
Manual operation: See "Log Message" on page 1164
:TEST:BB:BNC:SOURce <BncSource>
Selects the BNC connection test source.
This is a password-protected function. Unlock the protection level 1 to access it, see :
SYSTem:PROTect<ch>[:STATe].
Parameters:
<BncSource> AUTO | USER1 | USER2 | USER3 | USER4 | USER5 | USER6 |
TRGA | TRGB | C1TMC1 | C1TM2 | C1TM3 | C2TMC4 |
C2TM5 | C2TM6 | F1TMC1 | F1TM2 | F1TM3 | F2TMC4 |
F2TM5 | F2TM6 | F3TMC1 | F3TM2 | F3TM3 | F4TMC4 |
F4TM5 | F4TM6
*RST: USER1
Example:
SYSTem:PROTect1 0,123456
TEST:BB:BNC:SOURce AUTO
// detects the connectors that have assigned a signal and performs the connectror test
Manual operation: See "Source" on page 1164
TEST Subsystem