Remote Control Commands
R&S
®
SMW200A
1142User Manual 1175.6632.02 ─ 16
:TEST:BB:GENerator:FREQuency.................................................................................1144
:TEST:BB:GENerator:GAIN...........................................................................................1144
:TEST:BB:GENerator:SOURce......................................................................................1144
:TEST:BB:GENerator:STATe........................................................................................ 1145
:TEST<hw>:DIRect:BLOCk:READ?............................................................................... 1145
:TEST<hw>:DIRect:BLOCk:WRITe................................................................................1146
:TEST<hw>:ALL:STARt................................................................................................1146
:TEST<hw>:ALL:RESult?..............................................................................................1146
:TEST:BASeband?
Queries the result of the baseband selftest.
Return values:
<TestBbError> 0 | 1 | RUNning | STOPped
Example:
:TEST:BASeband?
// 0
// the test succeeded
Usage: Query only
Manual operation: See "Execute Baseband Selftest" on page 1167
:TEST:BB:BNC:CONNection?
Queries the BNC connection test result.
This is a password-protected function. Unlock the protection level 1 to access it, see :
SYSTem:PROTect<ch>[:STATe].
Return values:
<TestStatus> 0 | 1 | RUNning | STOPped
Example:
SYSTem:PROTect1 0,123456
TEST:BB:BNC:CONNection?
// RUN
// test is running
Usage: Query only
Manual operation: See "Execute BNC Connection Test" on page 1164
:TEST:BB:BNC:DESTination <BncDestination>
Selects the BNC connection test destination.
This is a password-protected function. Unlock the protection level 1 to access it, see :
SYSTem:PROTect<ch>[:STATe].