Configuring the Signal in the RF Domain
R&S
®
SMW200A
511User Manual 1175.6632.02 ─ 16
In the "Output" tab, you can configure the signal at the LF output, determine the
output voltage or add a DC offset.
State
Activates the output of the LF signal.
Remote command:
[:SOURce]:LFOutput<ch>[:STATe] on page 1066
Source
Determines the LF signal to be synchronized of the corresponding path.
"LF Generator A1/2, B1/2"
Selects one of internally generated LF signals.
"AM A/B"
Selects the AM signal.
"FM/ΦM A/B"
Selects the signal also used by the frequency or phase modulations.
"Noise Generator A/B"
Selects the internally generated noise signal.
"External 1/2"
Selects an externally supplied LF signal
Remote command:
[:SOURce]:LFOutput<ch>:SOURce on page 1066
Output Voltage
Sets the voltage (peak) of the LF output.
Remote command:
[:SOURce]:LFOutput<ch>:VOLTage on page 1067
DC-Offset
Adds a DC offset to the LF output signal.
Remote command:
[:SOURce]:LFOutput<ch>:OFFSet on page 1065
Show Connector
Accesses a dialog that displays the physical location of the selected connector on the
front/rear panel of the instrument.
A blinking LED on the front/rear panel indicates the selected connector, too (see "Con-
nector status indication" on page 663).
Analog Modulations