Remote Control Commands
R&S
®
SMW200A
1065User Manual 1175.6632.02 ─ 16
[:SOURce<hw>]:LFOutput:FREQuency:MODE <Mode>
Sets the mode for the output of the LF generator frequency, and determines the com-
mands to be used for frequency settings.
Parameters:
<Mode> CW | FIXed | SWEep
CW|FIXed
Sets the fixed-frequency mode. CW and FIXed are synonyms.
To set the output frequency, use command [:SOURce<hw>]:
LFOutput<ch>:FREQuency
SWEep
Sets sweep mode.
To set the frequency, use the commands:
[:SOURce<hw>]:LFOutput:FREQuency:STARt and [:
SOURce<hw>]:LFOutput:FREQuency:STOP
Or [:SOURce<hw>]:LFOutput:FREQuency:MANual
*RST:
CW
Example: See Example "Setup an LF sweep" on page 1061.
Manual operation: See "State (LF frequency sweep)" on page 476
[:SOURce<hw>]:LFOutput:FREQuency:STOP <Stop>
[:SOURce<hw>]:LFOutput:FREQuency:STARt <Start>
Sets the start/stop frequency for [:SOURce<hw>]:LFOutput:FREQuency:MODE
SWEep.
Parameters:
<Start> float
Range: 0.1 Hz to 1 MHz
Increment: 0.1
*RST: 1 KHz
<Stop> float
Range: 0.1 Hz to 1 MHz
Increment: 0.1 Hz
*RST: 100 KHz
Example: See Example "Setup an LF sweep" on page 1061.
Manual operation: See " Start Frequency/Stop Frequency " on page 478
[:SOURce]:LFOutput<ch>:OFFSet <Offset>
Sets a DC offset at the selected LF Output.